Co- 8.2.0 Isolate Very Intermittent Acid Pump Cycles; Co- 8.2.1 Aacid = Acid / Isolate Bicarb Pump Cycles; Co- 8.2.2 Abic = Bic - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 8.2.0 ISOLATE VERY INTERMITTENT ACID PUMP CYCLES

Call debug screen 0. Figure (right), WITHOUT LOOKING AWAY, watch Acid versus
AAcid for one (1) minute. Are they remaining always within +/- 3 of each other?
Yes
AAcid = Acid! See procedure number
No
AAcid is NOT remaining within 3 of Acid! Proceed to
PUMP
ERRORS...

CO- 8.2.1 AACID = ACID / ISOLATE BICARB PUMP CYCLES

Figure right, watch Bic versus ABic for one (1) minute. Are they remaining always
within +/- 3 of each other?
Yes
ABic = Bic! See procedure number
No.
ABic is NOT remaining within +/- 3 of Bic! Figure (right), watch the
Bicarbonate (BIC) pump's 'dot' for THREE (3) minutes. It should REMAIN
blue. TWO (2) possible scenarios:
1)
IF (and ONLY if) the 'dot' is EVER white: Proceed to
CO-
9.0.0.
2) IF the 'dot' REMAINS BLUE: Proceed to
PUMP
ERRORS.

CO- 8.2.2 ABIC = BIC

a) If not already, allow the external flow indicator's 'bob' to begin moving up and down
b) Based on the external meter's reading versus TCD, TWO (2) possible scenarios:
1) IF (and ONLY if) LESS THAN TCD - (minus) 0.4: See procedure number
2) IF MORE THAN TCD + 0.4 i.e.: See procedure number
CO- 8.2.1
CO- 8.2.2
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
(page 368).
page 397
,
SECTION 6 - CONCENTRATE
(page 368).
page 374
page 397
,
SECTION 6 - CONCENTRATE
page 369
CO- 8.3.0
(
, procedure number
CO- 8.3.0
(page 369).
).
Page 368