FIL- 5.1.0 FILACT = 0 / ISOLATE VENOUS PRESSURE (VEN)
Turn Dialysate Flow OFF i.e. Flow on/off lamp blinks!
a)
b) Call debug screen 0. ENSURE all eight balancing chamber valves (#31 through #38) 'dots' are
REMAINING white (Flow is Off)!
IMPORTANT! Dialysate Flow REMAINS OFF till instructed!
c)
Per the Figure below
d)
Level Detector's INNER P
e) Call debug screen 1. Push on the syringe plunger until VEN (middle column) is
between 400 and 450.
f)
Clamp the syringe tubing to HOLD the pressure!
g) If VEN falls more than 3 mmHg in 30 seconds there is a leak at the P
h) Leaving the clamp in place to keep VEN at more than 400, see procedure number
434).
, attach a syringe with a piece of tubing attached that will fit SNUG to the
port.
ven
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
port OR inside the module.
ven
FIL- 5.2.0
(page
Page 433