VEC- 1.2.0 'BOB' NOT MOVING / DIAVLO CYCLING BETWEEN 1 AND 0
a) Press 'Escape' to return to the Cleaning Program's Main screen!
b) Leave the machine in whatever Program it is CURRENTLY in: 1) In Rinse? OR 2) In Heat Disinfect?
1)
IF (and ONLY if) in RINSE: See procedure number
2) IF in HEAT DISINFECT: Per the Figure
right, does the VERY TOP Window,
Remaining Prerinse Time
min:sec (Yes or No)?
Yes
Remaining Prerinse Time =
0:00! See procedure number
VEC- 1.2.1
No
Remaining Prerinse Time does
NOT = 0:00! See procedure number
VEC- 1.3.0
VEC- 1.2.1 IN HEAT DISINFECT / REMAINING PRERINSE = 0:00
a) Call debug screen 0. If the screens do not appear press
'Escape' then call screen 0.
b) Only when Remaining Prerinse Time = 0:00, does Valve
#29's 'dot' (Figure right) REMAIN blue (i.e. NEVER white).
ENSURE this by watching it for fifteen (15) FULL seconds.
c) Figure right, screw a 60 ml syringe, filled COMPLETELY
with water, onto the Fluid Sample Port.
d) Start a SEVEN (7) second timer in your head as you
as HARD as you can
begin to push
syringe plunger! You should feel very little resistance!
e) Can you push ALL of the water through the Sample
Port within SEVEN seconds?
Yes Very little resistance is encountered! Valve
#29 is okay! See procedure number
VEC- 1.3.0 (page 209).
No Significant resistance when pushing! NOTE
ONLY VALVE #29 will be checked and
proceed to
TROUBLESHOOTING A VALVE
= 0:00
(page 208).
page 209
(
).
on the
page 192
,
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
VEC- 1.3.0
(page 209).
Page 208