Advertisement

CLEAN- 7.0.0 VALVE ERROR = 1 LONGER THAN TWO SECONDS

To avoid damage turn the machine OFF!
A)
B) Figure below, remove the distribution board cover.
C) FOUR (4) IMPORTANT checks!!
CHECK #1 ENSURE the ACTUATOR CABLE is plugged in SECURELY!
CHECK #2 Check the entire length of the ACTUATOR CABLE for damage!
CHECK #3 ENSURE all Valves are plugged PROPERLY into their distribution board positions!
CHECK #4 ENSURE the black PGND wire is plugged in and shows no signs of burning.
D) Was a problem located above?
Yes
Problem located! After the repair see procedure number
No problem located! Proceed to

CLEAN- 7.0.1 PROBLEM LOCATED

Return to Heat Disinfect OR Rinse!
a)
b) Call debug screen 0. If debug does not appear press 'Escape' then call screen 0.
c) Watch the 2
nd
window down,
two (2) seconds the repair solved the problem HOWEVER, if (and ONLY if) a Flow Error reoccurs see
(ABOVE) procedure number
page 190
, procedure number
Valve Error
, for five (5) minutes. If it
page 157
CLEAN- 1.0.5
(
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
CLEAN- 7.0.1
CLEAN-
7.3.0.
DOES NOT
).
(page 189).
= 1 for longer than
Page 189