Co- 10.0.00 Troubleshooting Online Clearance (Olc); Conductivity / Olc Components View - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 10.0.00 TROUBLESHOOTING ONLINE CLEARANCE (OLC)

OLC estimates UREA clearance (Kt/V) by measuring Conductivity pre and post dialyzer:
a) Return the dialyzer connectors to the shunt but
instructed!
Per the Figure below
b)
Cell #13) AND #44 (Post Temperature Sensor NTC #44), to the sensors to perform TWO (2) checks:
CHECK #1: Check the entire length for insulation damage. Damage will cause OLC problems and,
if damaged, the sensor should be replaced!
CHECK #2: ENSURE the connectors ARE NOT reverse connected another component! '!
c) From the Home screen, set [Dialysate Flow] to 500 ml/min and press 'CONFIRM'.
d) See procedure number
, trace the wires from distribution board positions #13 (Post Dialyzer Cond
CO- 10.0.10
(page 379).
Figure 62 – Conductivity / OLC Components
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
LEAVE THE DOOR OPEN till
Page 378