FIL- 2.0.6 ALARMS RESET / ISOLATE PRESSURE TEST
a) CLOSE the shunt door!
b) During this procedure you will look for air again through the INPUT TUBING at Chamber #69 but THIS
TIME while the Pressure Test's "Get Neg TMP" AND "Neg Flow On" banners are up as this is when air
may be pulled in through a leak somewhere!
c) RESET all alarms!
d) Select the [Test & Options] tab at the bottom of the screen. TWO (2) possible scenarios:
1)
IF the [Pressure Test] button is BLUE: A) Select [Pressure Test]; B) Press 'CONFIRM';
C) Allow the "Test: (Remove Air)" banner to expire THEN; D) IMMEDIATELY watch for air
bubbles at Chamber #69 for two (2) minutes!
2) IF the [Pressure Test] button is GRAY: A) Select [Both
Tests]; B) Press 'CONFIRM'; C) Allow the Conductivity Test to
complete; D) Allow the "Test (Remove Air)" banner to expire
THEN; E) IMMEDIATELY watch for air bubbles at Chamber #69 for two (2) minutes!
e) Are air bubbles seen MOVING into Chamber #69?
Yes
Air seen! Proceed to
No air seen! Allow the tests to finish then one (1) more minute BEFORE performing parts A through
H below:
A) Call the Home screen. Set [UF GOAL] to 1000 ml; Set [UF Time] to 1:00 hr.
B) Press 'CONFIRM'.
RESET ALL alarms!
C)
D) Turn the Blood Pump on so that it is rotating at more than 100 ml/min.
E) To avoid error, read parts F through H BEFORE performing them!
F) Press the front panel's UF on/off key. The green lamp MUST stay
on solid!
G) When TMP* reaches 280 turn UF off IMMEDIATELY (i.e. UF lamp
OFF)!
* If a "Filling Program" banner appears: 1) Repeat parts C) through G); 3) If "Filling
Program" reoccurs more than twice proceed to
FIL- 5.0.0
H) At Chamber #69's INPUT tubing, IMMEDIATELY watch for air again. Air seen?
Yes
No air seen! If you were prompted to Troubleshooting Filling Programs the original
page 432
, procedure number
page 432
Air seen! Proceed to
problem may still be occurring OTHERWISE do NOT continue!
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
FIL-
5.0.0.
page 432
, procedure number
, procedure number
FIL- 5.0.0