FIL- 2.0.0 VERIFY DIALYSATE SETTINGS
IMPORTANT! CLOSE THE SHUNT DOOR!
a)
b) Is the external flow indicator's 'bob' moving up and down in the sight tube?
Yes
'Bob' moving! See procedure number
No a) At the bottom of the screen, select the
'Dialysate' tab.
b) Figure right, if necessary adjust the
Conductivity Limits until 'Actual' Conductivity is
CENTERED between them.
IMPORTANT!
c)
d) Allow up to three (3) minutes for the
Conductivity window to turn white (i.e. No
Cond alarm).
e) Call debug screen 0. Figure right, allow Valve #24's
'dot' to turn BLUE BEFORE continuing to part f!
f)
Is the flow indicator's 'bob' moving up and down now?
Yes
No 'Bob' NOT moving!
the Home screen, [Dialysate Flow] is set to
800 ml/min! AND; 2) Screen 0's Valve #24's 'dot' is blue! If (and ONLY if) 'bob' is still
not moving proceed to
PROGRAM!
FIL- 2.0.1 'BOB' MOVING / ISOLATE POSSIBLE UF PUMP LEAK
a) Figure right, ENSURE the lamp above the front panel's UF on/off key
OFF
is
i.e. UF is OFF!
Parts b and c next page
Press 'CONFIRM'.
'Bob' moving!' See procedure number
FIL- 2.0.1
(page 416).
ENSURING:
page 20
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
FIL- 2.0.1
(page 416).
1) From
,
SECTION 1 - FLOW ERRORS IN DIALYSIS
Page 416