VE- 6.0.0 VALVE ERROR AT BALANCING CHAMBER CYCLE 2

Valves #31, #34, #36, and #37 'dots' REMAINING blue. One of these valve
circuits is causing the problem and these procedures isolate between them.
a) At the distribution board (Figure below), unplug
b) Call debug screen 0 and watch the Balancing Chamber Valve 'dots' for
forty-five (45) seconds. Do they begin to cycle between white and blue?
Yes
The Balancing Chamber Valves begin to cycle between white
and blue! Valve #31 OR its blue wire harness is bad.
No Valves DO NOT cycle! See procedure number

VE- 6.0.1 VALVE #31 UNPLUGGED AND VALVES DO NOT CYCLE

a) At the distribution board (Figure above), unplug
b) Watch the Balancing Chamber Valve 'dots' for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Yes
The Balancing Chamber Valves cycle between white and blue! Valve #34 OR its blue wire
harness is bad.
No Valves DO NOT cycle! See procedure number

VE- 6.0.2 VALVE #34 UNPLUGGED AND VALVES DO NOT CYCLE

a) At the distribution board (Figure above), unplug
b) Watch the Balancing Chamber Valve 'dots' for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Valve #31's
Figure 39 – Valve Cycle 2
Valve #34's
Valve #36's
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
connector.
VE- 6.0.1
(page 203).
connector.
VE- 6.0.2
(page 203).
connector.
Page 203