F- 9.0.41 LESS THAN 800 ML TO VALVE #26

a) ENSURE no restrictions or leaks between the DiaSafe
b) ENSURE no restrictions between the DiaSafe
Figure below
c)
, at the rear of the DiaSafe
direct it into the bucket.
d) Return to Calibrate Hydraulics → Flow Pressure.
e) Press 'CONFIRM' then ENSURE the Flow Motor is rotating!
f)
Direct the tubing into the graduated cylinder for one (1) minute.
g) Press 'CONFIRM' to return to "Calibrate Hydraulics".
h) More than 800 ml collected?
Yes
More than 800 ml collected! The DiaSafe
No
Less than 800 ml collected! This indicates a restricted 'Fresh Side' Balancing Chamber valve.
See parts THROUGH c below:
a) Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program ("Select Program" → 'Dialysis' →
'CONFIRM').
c) Proceed to
®
filter and Valve #26.
®
filter housing, remove the tubing from Valve #28 (V28) and
page 62
procedure number F- 7.0.0
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
®
filter and the Balancing Chambers.
®
filter is restricted!
Page 98