F- 9.0.4 INPUT FLOW PUMP FLOW LESS THAN 800 ML / ISOLATE V26 INPUT FLOW
Remove the clamps from Valve #43!
a)
b) Obtain an empty bucket.
Figure below
c)
, remove Valve #26's INPUT tubing and direct it into the bucket!
d) Return to Calibrate Hydraulics → Flow Pressure.
e) Press 'CONFIRM' then ENSURE the Flow Motor is rotating!
f)
Direct Valve #26's tubing into the graduated cylinder for one (1) minute.
g) Press 'CONFIRM' to return to "Calibrate Hydraulics"!
h) Reattach Valve #26's tubing.
i)
MORE than 800 ml collected?
Yes
MORE than 800 ml! Proceed to
No
Less than 800 ml! See procedure number
page 99
F- 9.0.41
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number
F-
9.0.42.
(page 98).