E) Per the Figure below AND as directed in
Sensor Board's X2 connector starting at pin #2:
Figure 90 – Electronic Card Cage (Sensor Board)
Table 10 – Sensor Board 'X2' DC Voltage Checks (Refer also to
X2 Pin
X2 Pin Location
2
Top row, first pin
from the REAR of
the machine
6
Top row, third pin
from the REAR of the
machine
7
Bottom row, fourth
pin from the REAR of
the machine
Table 11
below, measure from the solder (back) side of the
• IF less than 0.5 volts DC: Measure at pin 6 (next row in table)
• IF more than 0.5 volts DC: Bad Actuator-Test Board (see page 9).
• IF less than 4.5 volts DC: See procedure number
• IF more than 5.2 volts DC: See procedure number
• IF between 4.5 and 5.2 volts DC: From debug screen 4, does LEAK = the
pin 6 measurement within +/- 0.3 volts DC?
Yes
LEAK = pin 6 (+/- 0.3). Measure at pin 7 (next row in table).
No
LEAK NOT within +/- 0.3 of pin 6 = Bad Sensor Board (see page 9).
• IF less than 4.0 volts DC: See procedure number
• IF more than 6.0 volts DC: See procedure number
• IF between 4.0 and 6.0 volts DC: From debug screen 4, does DIMN = the
pin 7 measurement within +/- 0.3 volts DC?
Yes DIMN = pin 7 (+/- 0.3). See procedure number
No DIMN does NOT within +/- 0.3 of pin 7= Bad Sensor Board (see page 9).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Figure
Your Response:
BL- 6.0.0
BL- 6.0.0
BL- 6.0.0
BL- 6.0.0
BL- 6.0.0
90)
(page 520).
(page 520).
(page 520).
(page 520).
(page 520).
Page 519