BL- 3.0.5 AFTER INDUCED POSITIVE PRESSURE TESTS
a) Other procedures, in different Sections of the Guide, are performed next
page and procedure number (BL- 3.0.5) because you may prompted to return to here
b) Before continuing to part c, proceed to
PRESSURE TESTS
c) If no leaks were located in part b return to Dialysis Program ("Select Program" → 'Dialysis' →
'CONFIRM')!
d) ALLOW six (6) minutes for [Temperature] AND [Conductivity] to STABLIZE to normal.
e) ENSURING the external flow indicator is moving up and down, press and release the 'RESET' key up to
three (3) times in twenty (20) second intervals attempting to clear a Blood Leak alarm BEFORE
continuing to part f.
f)
Once again, watch for air bubbles through the Blood Leak Sensor. THREE (3) possible scenarios 1) or 2)
or 3) below.
1)
IF (and ONLY if) the Blood Leak alarm will RESET AND air is still seen: Proceed
490
,
SECTION 11 - INDUCED AIR LEAK
2) IF (and ONLY if) the Blood Leak alarm will NOT RESET AND air is still seen: Locate and repair
the air leak!
3) IF no air is seen: See procedure number
page 496
, to perform
TESTS.
BL- 4.0.0
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
. IMPORTANT!
SECTION 12 - INDUCED POSITIVE
(page 516).
NOTE this
to page
Page 515