FIL- 3.6.2 FILACT = 1 / CHECK FOR AIR AT CHAMBER #69
Figure below, watching for two (2) minutes, are air bubbles seen MOVING INTO Air Removal Chamber #69
now?
Yes
Air seen! Proceed to
source of the air leak!
No
No air seen! Watch FILACT again for one (1) minute or until it EVER = 1! THREE (3) possible
scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) FILACT always = 1 (NEVER = 0): See procedure number
(page 430).
2)
IF (and ONLY if) FILACT now REMAINS = 0: Proceed to
FIL- 4.0.0
3)
IF FILACT cycles between 0 and 1: A procedure, in different Section, is performed next.
NOTE this page and procedure number (FIL- 3.6.2) as you may prompted to return here.
Perform parts a THROUGH c below:
a) BEFORE continuing to part a, proceed to
DIAGNOSTIC VALVE LEAK
b) If a leaking Balancing Chamber valve was not located in part a, return to Dialysis
Program and wait until the external flow indicator's bob moves up and down (i.e.
good Temp and Cond).
c) Call debug screen 1. Watch FILACT again for one (1) minute or until it EVER = 1.
TWO (2) possible scenarios:
1)
2) IF FILACT now REMAINS = 0: Proceed to
page 490
,
SECTION 11 - INDUCED AIR LEAK TESTS
IF (and ONLY if) FILACT EVER = 1: See procedure number
430).
FIL- 4.0.0
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 431
page 527
TESTS.
page 431
to locate the
FIL- 3.7.0
, procedure number
to perform
SECTION 18A –
FIL- 3.7.0