Tmp- 1.0.5 Loading Pressure Okay/Isolate Air Removal System - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

TMP- 1.0.5 LOADING PRESSURE OKAY / ISOLATE AIR REMOVAL SYSTEM
Turn the water OFF!
a)
b) The Deaeration gauge is used next. ENSURE it reads 0
inHg before installing it!
c) Allow a "No Water" alarm to occur THEN, after one (1)
minute, the deaeration motor stops.
d) Figure right, tee the gauge into the Inlet (clear tubing)
side of the Deaeration Pump.
IMPORTANT! Turn the water on and allow
e)
the "No Water" alarm to go away!
f)
Is Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes
Deaeration pressure OKAY! See procedure number
No
Deaeration Pressure in NOT OKAY! ENSURING a "No Water" alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
- DEAERATION
Appendix A
PROBLEMS.
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
TMP- 2.0.0
(page 446).
page 498
,
SECTION 13
Page 445