CO- 4.2.3 VERIFY CONDUCTIVITY
a) From the Home screen, set [Dialysate Flow] to 800 ml/min.
b) Allow six (6) minutes BEFORE continuing to part c.
c) Does [Conductivity] REMAIN less than 13.0 mS?
Yes
REMAINS LESS than 13.0 mS! Return to
No
Conductivity MORE than 13.0 mS! The low Conductivity problem is solved.
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 325
, procedure number
CO-
1.0.4.
Page 347