F- 7.4.0 VERIFY DRAIN FLOW
The "Select Program" screen remains up till instructed otherwise!
a)
b) Was there ANY drain flow at all?
Yes
There was some flow! Proceed to
No
There was no flow at all! See procedure number
F- 7.4.2 NO DRAIN FLOW
Figure below, EXCEPT for valves #38 and #31, ENSURE are ALL OTHER valves, especially Valve #30,
are plugged PROPERLY into their distribution board positions?
Yes
Valves plugged in correctly! See procedure number
No
You located a valve plugged wrong! Return ALL VALVES to their distribution board PROPER
positions and return to
page 79
page 76
, procedure number
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number
F- 7.4.3
F- 7.4.2
(page 78).
F- 7.4.3
(page 79).
F-
7.3.2.
Page 78