Tmp- 2.0.7 No Uf Pump Leaks - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes
The UF 'dot' blinks blue! If SURE the green UF lamp was OFF, the Actuator-Test Board may
be bad!
No The 'dot' REMAINS white!
tubing! Look at the UF Pump's nozzle! TWO (2) possible scenarios:
1)
IF (and ONLY if) NO fluid output from the nozzle! A) Reattach the tubing! B) See
procedure number
2)
IF fluid output from the nozzle! TWO (2) possibilities: 1) The UF Pump is upside
down! OR; 2) Incorrectly installed OR bad UF Pump seals/springs.

TMP- 2.0.7 NO UF PUMP LEAKS

IMPORTANT! Remove the 'dummy chamber' from the Level Detector module!
a)
IMPORTANT! CLOSE the shunt door!
b)
c)
From debug screen 0, Figure right, allow Valve #24's 'dot' to
REMAIN BLUE i.e. normal AND stable Temp, Cond AND shunt
door closed!
d) The external flow indicator's 'bob' MUST be moving up and
down!
e) Call debug screen 1. Watch FILACT for thirty (30) seconds to
ENSURE it = 0 ALWAYS.
f)
See procedure number
Per
Figure 70
TMP- 2.0.7
Figure 70 – UF Pump OUTPUT
TMP- 2.0.8
(page 454).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
below
, remove the UF Pump's output (TOP)
(page 453).
Page 453