Fresenius Medical Care 2008K2 Troubleshooting Manual page 338

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 1.0.8.5 AFTER RINSE CONDUCTIVITY = 10.0 MS
a) Select the [Dialysate Flow] window.
b) Set [Dialysate Flow] to 800 ml/min and press 'CONFIRM'.
page 337
c) Allow six (6) minutes BEFORE continuing to procedure number CO- 2.0.0 (
).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
Page 334
P/N 507298 Rev. H