Cable Plug / Sensor Board X2 Pin Locations Table - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

i)
Per the Figure below, the cable terminates at the Sensor Board's X2 connector. The solder (rear) side
PINS are in two rows. Odd numbered in the bottom, even numbered in the top. NOTING the reference
to the 'Rear of Machine' locate the appropriate PIN number per
j)
Place the second meter on the matching numbered PIN on the solder (rear) side of the P2 connector.
Some components require checking multiple conductors. 2.0
conductors?
Yes
The cable is good! Return it to the distribution board. If referred to CHECKING THE
SENSOR BOARD CABLE, return to the procedure that brought you here as noted.
No
Unplug the cable from the Sensor Board's X2 connector. Check inside the connector for
damaged male pins. If the pins are okay, replace the cable.
Figure 94 – Electronic Card Cage (Sensor Board)
Table 13 – Cable Plug / Sensor Board X2 Pin Locations
HYDRAULIC COMPONENT
NTC #3
NTC #2
FLOAT
CFS TRANSDUCER (1 of 3)
CFS TRANSDUCER (2 of 3)
CFS TRANSDUCER (3 of 3)
COND CELL
ACID EOS
Table 14 continued next page
Cable Plug
/ X2 PIN #
12
13
21
11
14
15
16
25
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Table 14
(page 525).
Ω
or less for ALL required
Cable Plug / X2 PIN LOCATION
Top row, 6 pins from the rear of machine
Bottom row, 7 pins from the rear of machine
Bottom row, 7 pins from the front of machine
Bottom row, 6 pins from the rear of machine
Top row, 7 pins from the rear of machine
Bottom row, 8 pins from the rear of machine
Top row, 8 pins from the rear of machine
Bottom row, 5 pins from the front of machine
Page 525