MOTORS- 4.0.0 MOTOR NOT ROTATING

NOTED
Hit the
motor HARD with a screwdriver handle. Does it start rotating?
Yes
Motor rotating! Proceed to
No Motor does NOT start rotating! Perform parts a THROUGH e below:
To prevent damage TURN THE MACHINE OFF!
a)
b) Figure below, remove the distribution board cover.
c) TWO (2) checks:
CHECK #1: ENSURE the pump is plugged in PROPERLY!
CHECK #2: ENSURE the Actuator Ribbon Cable is plugged in securely!
d) Turn the machine ON.
e) Depending on what Program the machine was in when you ORIGINALLY started
troubleshooting:
1)
IF a Cleaning Program (Heat Disinfect, Rinse, etc.): Place the machine in RINSE then
see procedure number
2)
IF Dialysis Program: Perform parts A through C below:
A) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM').
B) From the Home screen, set [Dialysate Flow] to 800 ml/min.
C) See procedure number
page 138
, procedure number
MOTORS- 4.1.0
MOTORS- 4.1.0
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
MOTORS-
5.0.0.
(page 134).
(page 134).
Page 133