Fresenius Medical Care 2008K2 Troubleshooting Manual page 331

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

FIGURE BELOW
B)
Sensor's distribution board position i.e.
NOTE:
*
If CBE modified the resistor MUST into the CBE board pin for pin! ENSURE the top CBE
board pin is covered by the resistor!
C) If the resistor plug was placed properly, watching to thirty (30) seconds,
Chamber #69's box MUST now say "No Air" always!
D) See procedure number
, place one of the plugs, from the
th
5
position
CO- 1.0.6
(page 328).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
FOUR-RESISTOR SET
* from LEFT
into the Air
Page 327