SECTION 28 – HEAT EXCHANGER LEAKING EXTERNALLY
A) Per the Figure below, if the HEAT EXCHANGER'S 'center screws' and nuts are not tight this may be
problem!
B) A psi pressure gauge is required! ENSURE it reads 0 psi before installing it!
IMPORTANT!
C)
Turn the water OFF.
D) Per the Figure below, install the gauge at Inlet Pressure Regulator #61.
IMPORTANT!
E)
Turn the water on!
F) ENSURE NO LEAKS at the gauge!
G) Plug the concentrate connectors into their rinse ports.
H) Place the machine in
Part I next page
Figure 135 – Inlet Water Regulator #61
RINSE
!
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 665