CO- 1.0.10 VERIFY ACID SETTINGS
a) Select the upper left 'Conc' button to call the ACID Sub Screen (Figure below). This screen indicates
the 'Selected Acid's' Ionic Profile.
b) Does the label on the connected ACID match the Sub Screen's Ionic Profile (Yes or No)?
Yes
The label = the screen's Ionic Profile. See parts a AND b below:
a) Per its label, ENSURE the bicarb is the same 'Type' as the acid.
b) See procedure number
No Referring to NOTES 1 through 4 below, THREE (3) possibilities: 1) The connected acid may
NOT be the 'Selected Acid'; 2) You made an error and the levels are correct; 3) The connected
acid may have been entered incorrectly in Service mode.
NOTE 1:
If a Granuflo
ACID'S label. EXAMPLE: 4.0 on the screen = 8.0 on the label. If Granuflo
the screen's Ac value should = exactly the value on the label!
NOTE 2:
Sub Screen values are rounded! EXAMPLE: 2.30 on the screen may = 2.25 on the label.
NOTE 3:
Na+ = Sodium. If the sodium level on the label is 110 or less it indicates sodium contributed
by the acid alone. If more than 110 it indicates the TOTAL or FINAL Sodium.
NOTE 4:
K+ = Potassium; Ca++ = Calcium; Mg++ = Magnesium; Ac = Acetate; Citrate is NOT
ALWAYS present; Dex = Dextrose (mg/dL*).
* If the label states Dextrose in mmol/L THEN 5.5 mmol/L = 100 mg/dL;11 mmol/L = 200
mg/dL
ACID Sub Screen
CO- 2.0.0
®
acid is selected the screen's Acetate (Ac) value should be ½ the value on the
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 337
(
).
®
IS NOT selected
Page 336