Ve- 1.0.0 Initialize Actuator-Test Board Valve Cycles - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

A resistor plug IS NOT ALREADY in place continued:
a) Unplug the female Air Sensor's connector,
position, from the LEFT!
b) Call debug screen 0. Figure right, the text box above Chamber #69
MUST say "Air"!
c) Place one of the plugs, from the
into the Air Sensor's distribution board position!
* If CBE modified the resistor plugs into the CBE board pin for pin. ENSURE the TOP CBE
board pin is covered by the resistor
d) If the resistor plug was placed properly, Chamber #69 MUST
now say "No Air"!
e) See procedure number

VE- 1.0.0 INITIALIZE ACTUATOR-TEST BOARD VALVE CYCLES

Turn the machine OFF!
a)
b) Turn the machine ON and return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!
c) From the Home screen, set [Dialysate Flow] to 300 ml/min.
d) See procedure number
FOUR-RESISTOR SET
VE- 1.0.0
(page 196).
VE- 1.0.2
(page 197).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H