Clean- 4.0.0 Good Flow Pump Pressure/Isolate The Cleaning Program; Clean- 4.2.0 Remaining Prerinse Time - 0:00 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CLEAN- 4.0.0 GOOD FLOW PUMP PRESSURE / ISOLATE THE CLEANING PROGRAM
a) Leave the gauge installed
Open the shunt door and LEAVE IT OPEN until instructed!!
b)
c) Leaving the machine in whatever Program it is CURRENTLY in: 1) Rinse? OR 2) Heat Disinfect?
1)
IF (and ONLY if) in RINSE: Proceed to
2) IF in HEAT DISINFECT: See parts a AND b below:
a) Press 'Escape' to call the Heat
Disinfect screen (Figure right).
b) Ignoring the Remaining Time
window, does the VERY TOP
Remaining Prerinse
window,
Time
, = 0:00 min:sec?
Yes
Remaining Prerinse Time = 0:00!
See procedure number
CLEAN- 4.2.0
No Remaining Prerinse Time does
NOT = 0:00! Proceed to
184
, procedure number
CLEAN- 5.0.0
CLEAN- 4.2.0 REMAINING PRERINSE TIME = 0:00
a) Call debug screen 0. If the debug screens do not appear
press 'Escape' then call screen 0.
b) When Remaining Prerinse Time = 0:00 Valve #29's 'dot'
(Figure right) REMAINS blue. WITHOUT LOOKING AWAY,
watch it for twenty (20) FULL seconds.
c) TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve #29's 'dot' is ALWAYS blue:
See procedure number
2) IF Valve #29's 'dot' is EVER = white: Proceed to
ENSURE
its tubing is NOT kinked (restricting flow)!
page 184
(page 175).
page
.
CLEAN- 4.3.0
(page 176).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number CLEAN- 5.0.0.
page 184
, procedure number CLEAN- 5.0.0.
Page 175

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2008k

Table of Contents