Co- 9.0.0 Bic Dot Is White/Isolate Bicarb Debug Signal - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 9.0.0 BIC DOT IS WHITE / ISOLATE BICARB DEBUG SIGNAL
A) Per the Figure below, inside the distribution board, unplug the
This is the Bicarb Reed Switches Connector, "X12, BIC-SW".
B) Using a flashlight, check inside the "BIC-SW" position for corrosion or damaged male pins as this may
be the problem!
C) Call debug screen 0.
D) WITHOUT LOOKING AWAY, watch the bicarbonate (BIC) pump's 'dot'
(Figure right) for two (2) minutes. Does it REMAIN AWAYS blue?
Yes
Bic 'dot' remains blue! Referring to the
Figure right, the Bicarb Reed Switch
(#12) is bad.
No
Bic 'dot' remains white OR cycles between white
and blue! Continue to part E.
E) Leave the BIC Reed Switch unplugged till the problem
is solved i.e. until the Bic 'dot' stays blue!
F) Turn the machine OFF!
G) One at a time, swap in the listed components (see
COMPONENT LIST below) with known good THEN
perform parts H and I to test the new component.
COMPONENT LIST: 1) Actuator-Test Board; 2) Sensor Board*; 3) Sensor Board cable; 4) Functional
Board*; 5) Distribution board; 6) Motherboard.
* For each board, to prevent "COND OFFSET" alarms place the machine into T and C Mode (refer to
OPERATING
MODES, page Error! Bookmark not defined.)
H) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM').
I)
Call debug screen 0. If the Bic 'dot' REMAINS blue the last component swapped in was the problem. If
still white or cycling between white and blue return the to part F to swap in the next component
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
th
9
CONNECTOR CAP from the left!
Page 374

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2008k

Table of Contents