T- 1.5.4 [TEMPERATURE] DOES NOT = 95° C BUT "TEMP OVER 95 DEGREES" REOCCURED
A) Read this step before performing it!
Temp Sensor
Calibrate
Procedures booklet but does "Operator Error"
OR "Actuator Board Error" occur during the
calibration?
Yes
An "Error" banner occurs! CAREFULLY repeat the
"Error" banner reoccurs proceed to
DOES NOT reoccur see part B.
No
An "Error" banner DOES NOT occur! See part B.
B) Read before performing! Return to
fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.4 continue to part C.
C) BEFORE continuing to part D calibrate
D) Read before performing! Return to
fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.4 continue to part E.
E) BEFORE continuing to part F turn the machine off and swap in a known good Actuator-Test Board*. *To
LOCATE the board refer to
F) BEFORE continuing to part G, note this page number then proceed to
VOLTAGE DETECTOR CALIBRATION
G) Read before performing! Return to
fixed the problem. If (and ONLY if) you return to procedure number T- 1.5.4 replace the Functional
Board* and perform all calibrations. *To LOCATE the board refer to
per the Calibration
page 302
page 247
, procedure number
Temp Control
page 247
, procedure number
Figure 4A
(page 9).
page 247
, procedure number
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Temp Sensor
calibration but if an
, procedure number
T-
7.0.0. If an Error banner
T- 1.5.1
to see if the calibration
per the Calibration Procedures booklet.
T- 1.5.1
to see if the calibration
page 14
T- 1.5.1
to see if the new board
Figure 4A
(page 9)
to perform the
Page 250