BL- 1.0.5 BLOOD LEAK ALARM?

Is the Blood Leak alarm still present (Yes or No)?
Yes
Press and release the 'RESET' key up to three (3) times in twenty (20) second intervals. TWO (2)
possible scenarios:
1)
IF the Blood Leak alarm clears: See procedure number
2) IF the Blood Leak alarm does NOT clear: See procedure number
No The alarm clears! See procedure number

BL- 1.0.6 BLOOD LEAK ALARM NOT PRESENT (CHECK FOR AIR)

a) RESET ALL alarms and turn the blood pump on to 100 ml/min or more. NOTE: If the blood pump will not
run perform the alarms tests.
b) From the Home screen, set [UF GOAL] to 1000 ml; set [UF Time] to 1:00 hr.
c) Press 'CONFIRM'.
d) To turn UF on, press the front panel's UF on/off key. The green on/off lamp
turns on solid.
e) Allow TMP to increase to 160 mmHg THEN IMMEDIATELY turn UF off (UF lamp off). TMP should
increase to no more than 300 mmHg
1
If a "Filling Program" banner appears allow it to clear. If "Fillling Program" occurs more than twice an air
leak may be indicated. In this event (ONLY) proceed to
f)
Per the Figure right,
flashlight
, watch for air bubbles
through the Blood Leak Sensor's TO
AND FROM tubing for one (1) minute
each.
g) TWO (2) possible scenarios:
1)
IF (and ONLY if) air is seen! See
procedure number
page 514
(
).
2)
IF no air is seen: Proceed to
page 520
, procedure number
BL- 6.0.0
1
.
using a
BL- 3.0.0
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
BL- 1.0.6
(page 512).
page 514
BL- 1.0.6
(page 512).
page 513
BL- 2.0.0
(
, procedure number
BL- 3.0.0
).
Page 512