Co- 10.0.60 Isolate Post Sensors Stability - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 10.0.60 ISOLATE POST SENSORS STABILITY

a) Per the Figure below, remove Post Dialyzer Temp Sensor NTC #44's connector from distribution board
position "x44, NTC-POST".
b) Using a flashlight, check the vacant "x44" position for 'white' corrosion and/or damaged male pins. If
damage is located this may be the problem!
c) Per the Figure right, place the
into the Post Dialyzer Temp Sensor's vacant distribution board position "x44,
NTC-POST".
d) Per the Figure above, remove Post Dialyzer Cond Cell #13's connector from
distribution board position "x13, COND-POS".
e) Check the vacant "x13" position for 'white' corrosion and/or damaged male pins.
274 Ω
f)
Place the
resistor plug, from the TWO-RESISTOR SET, into the Post Dialyzer Cond Cell's #13
distribution board position "x13, COND-POS".
g) Call debug screen 5. TWO (2) checks then proceed according to the TWO (2) possible scenarios after
check #2:
CHECK #1 (TPOS): WITHOUT LOOKING AWAY, watch TPOS (4
right, at top) for exactly one (1) minute noting its highest and lowest values. Subtract
the lowest value seen from the highest. 2 (two) or less = GOOD! MORE THAN 2 =
BAD!
CHECK #2 (CPOS): Now watch CPOS (4
exactly one (1) minute noting its highest and lowest values. Subtract the lowest
value seen from the highest. 8 (eight) or less = GOOD! MORE THAN 8 is BAD!
IF (and ONLY if) TPOS AND / OR CPOS ARE BAD:
1)
procedure number
IF TPOS AND CPOS are both GOOD:
2)
a) Return Post Dialyzer Temp Sensor NTC #44's connector to distribution board position labeled
"x44, NTC-POST".
Parts b through d next page
6.04 KΩ
plug, from the TWO-RESISTOR SET
th
column from right, third box down) for
CO-
11.0.00.
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
th
column from
Proceed to
See parts a through d below:
page 394
,
Page 391