CO- 2.0.8.1 VALVE #24'S 'DOT' IS BLUE
Is the external flow indicator's 'bob' rising at least ¼ way up in the sight tube?
Yes
'Bob' moving! See procedure number
No
'Bob' NOT moving. See parts a THROUGH c below:
a) From the Home screen, ENSURE [Dialysate Flow] remains set to at least 500 ml/min!
b) ENSURE the [Temperature] AND [Conductivity] windows are pale yellow/white.
c) If 'bob' STILL is not moving AND if (and ONLY if) debug screen 0's, Valve #24's 'dot' is still
blue, proceed to
CO- 2.0.9
page 20
,
SECTION 1 - FLOW ERRORS IN DIALYSIS
LEFT BLANK INTENTIONALLY