F- 6.4.0 Valve #30 Is Open; F- 6.4.2 Isolate Valve #30 Output Flow - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

1)
IF (and ONLY if) the plunger moves with NO resistance: Valve #30 is open! See procedure
number
F- 6.4.0
2) IF the plunger OFFERS RESISTANCE: Valve #30 is restricted! TWO (2) possible bad
components: 1) Bad Valve #30 OR; 2) Bad Actuator-Test Board. To LOCATE the board refer to
Figure 4A
(page 9).

F- 6.4.0 VALVE #30 IS OPEN

a) Reconnect valve #30's INPUT tubing but DO NOT reattach the drain tubing to the rear the machine!
b) Obtain a 1000 ml graduated cylinder to measure flow subsequently.
c) Place a bucket at the drain nozzle to capture subsequent potential flow.
d) Turn Dialysate Flow ON (Flow on/off lamp NOT blinking).
e) From the Home screen, ENSURE [Dialysate Flow] is set to 800 ml/min!
f)
DO NOT turn flow off again to ENSURE measurement accuracy below!
g) TWO (2) possible TMP window scenarios:
1)
IF (and ONLY if) the TMP window is NOT red: See procedure number
2)
IF the TMP window is red (TMP alarm): See parts A THROUGH C below:
A) Press and release the 'Reset' key then immediately press and hold it for three (3) seconds.
B) Allow thirty (30) seconds.
C) If a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to procedure
number
F- 6.4.2

F- 6.4.2 ISOLATE VALVE #30 OUTPUT FLOW

a) ALLOW one (1) minute BEFORE continuing to part b!
b) Measure flow from the REAR 'To Drain' port for one (1) minute. TWO (2) possible scenarios:
1)
IF (and ONLY if) LESS THAN 725 ml: Proceed to
F-
7.0.0.
2) IF MORE THAN 725 ml: Check the drain tubing and it's (if used) the'Quick Connector' for possible
bio-growth restrictions.
(page 60).
(page 60).
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
F- 6.4.2
page 62
, procedure number
(page 60)
Page 60