Tmp- 3.4.0 Acfs Not Between 3.5 And 6.0/Isolate Pdia; Tmp- 3.5.0 Pdia Less Than 4.0/Isolate Transducer #9 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

TMP- 3.4.0 ACFS NOT BETWEEN 3.5 AND 6.0 / ISOLATE PDIA
From debug screen 10, based on PDIA (middle column). TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA = less than 4.0: See procedure number
2) IF PDIA = more than 6.0: Proceed to
TMP- 3.5.0 PDIA LESS THAN 4.0 / ISOLATE TRANSDUCER #9
a) Figure right, unplug the Dialysate Pressure Transducer
#9 from distribution board position X9, "P-DIAL".
IMPORTANT!
b)
To avoid unnecessary work, using
the screen's clock (upper right), allow up to five (5)
minutes (PDIA response is NOT instantaneous).
PDIA should increase to 9.0 or more?
Yes
PDIA increased to 9.0 or more! See procedure number
No
PDIA DOES NOT increase to 9.0 or more! See parts A through D below:
A) Turn the machine OFF.
B) Swap in the following components (COMPONENT LIST below), one at a time, with known
good, and in between perform parts C and D to see if the new component fixed this
problem.
COMPONENT LIST: 1) Actuator-Test Board
4) Functional Board
1
To locate the boards refer to
2
To prevent "Cond Offset Failure", place the machine into T and C Mode (refer to
OPERATING MODES
C) Leaving distribution board position "x9, "P-DIAL" VACANT return to Dialysis Program.
D) If screen 10's PDIA is still NOT more than 9.0 return to part A!
page 465
, procedure number
1, 2
; 5) Distribution board; 6) Motherboard
Figure 4A
(page Error! Bookmark not defined.)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
TMP- 3.5.0
TMP-
TMP- 3.6.0
2
1
; 2) Sensor Board
1
(page 9)
(page 464).
3.6.0.
(page 465).
; 3) Sensor Board cable;
Page 464