Co- 2.0.0 Verify Primary Temperature Alarm Limits; Co- 2.0.6 Verify Primary Conductivity Alarm Limits; Co- 2.0.61 [Conductivity] Window Is Red - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

CO- 2.0.0 VERIFY PRIMARY TEMPERATURE ALARM LIMITS

a) Call the Home screen.
b) ENSURING [Temperature] REMAINS between 35.5 and 38.5° C,
what color is its window (Figure right)?
1)
IF pale yellow/white: Proceed to
2)
IF RED: Proceed to

CO- 2.0.6 VERIFY PRIMARY CONDUCTIVITY ALARM LIMITS

Call the Home screen. What color is the [Conductivity] window (Figure
right)? TWO (2) possible scenarios:
1)
IF (and ONLY if) pale yellow / white: See procedure number
2) IF RED: See parts a AND b below:
a) At the bottom of the screen, select the 'Dialysate' tab.
b) Figure right, using [Alarm Position], can the Limits be
adjusted so that 'Actual' Conductivity is
CENTERED between them?
Yes
'Actual' Conductivity between the Limits!
Proceed to
CO-
2.0.7.
No Limits cannot be adjusted properly! See procedure number

CO- 2.0.61 [CONDUCTIVITY] WINDOW IS RED

a) Figure right, place the
resistor plug, from the TWO-
RESISTOR SET, into distribution
board position, "X7, COND".
b) Can the Limits now be adjusted so
that 'Actual' Conductivity is
between them?
Yes
'Actual' Conductivity between the Limits! See procedure number
No The Limits CANNOT be adjusted properly! See parts a THROUGH d below:
a) Return Conductivity Cell #7's connector to distribution board position "X7, COND".
b) Return the concentrate connectors and close the shunt door.
Parts c and d next page
page 337
page 222
, procedure number
page 339
, procedure number
274 Ω
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number
CO-
T- 1.0.9
.
CO- 2.0.8
(
CO- 2.0.61
2.0.6.
page 340
).
(page 337).
CO- 2.0.7
(page 339).
Page 337