CLEAN- 1.2.3 ISOLATE LOADING PRESSURE (3)

A)
Loading Pressure may or may not be cycling. TWO (2) possible scenarios below:
1)
IF NOT cycling: Pressure is 'OKAY' if it REMAINS between 22 and 27 psi. See part B!
2)
IF cycling: It may cycle to but SHOULD NOT remain less than 15 psi. It is 'OKAY' if it EVER
cycles to a PEAK of somewhere between 22 and 27 psi. See part B!
B)
FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) Loading Pressure is 'OKAY': Leaving the gauge installed, proceed
172
, procedure number
2)
IF (and ONLY if) REMAINS ALWAYS less than 15 psi: ENSURING the gauge was SLAMMED
into the Rinse port, Loading Pressure is low! See procedure number
3)
IF (and ONLY if) EVER MORE than 28 psi: Turn Valve #65's nut (Figure below)
counterclockwise (outward) to adjust to a PEAK of between 22 and 25 psi. If a Flow Error
reoccurs proceed
4)
IF cycling to a PEAK of between 15 and 21 psi: ENSURING the gauge was SLAMMED into
the Rinse port, turn Valve #65's nut (Figure below) clockwise (inward) attempting to adjust to a
PEAK of between 22 and 25 psi. Can you?
Yes
No Loading Pressure remains low! See procedure number
CLEAN- 3.0.0
page 172
, procedure number
Pressure CAN be adjusted to its PEAK! Allow two (2) minutes then if a Flow Error
page 172
reoccurs proceed
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
CLEAN-
3.0.0.
, procedure number
CLEAN-
CLEAN- 1.2.33
page
CLEAN- 1.2.33
(page 161).
3.0.0.
(page 161).