Fresenius Medical Care 2008K2 Troubleshooting Manual page 39

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

a) Momentarily plug JUST the acid (red) connector into its rinse port to call "Select Program"!
b) RETURN the acid connector to concentrate!
c) Select the 'Dialysis' button but
d) Call debug screen 0. If parts a through c were performed correctly the TOP balancing chamber
valves 'dots', #31 through #34, are REMAINING BLUE!
e) Remove the red DIALYZER connector from the shunt door
and HOLD IT UP as seen in the Figure right!
f)
Continuous flow, more than 0.2 ml per minute from the connector?
Yes
Continuous flow! Turn the machine off THEN proceed to
18A – DIAGNOSTIC VALVE LEAK
No
No Flow! See parts a AND b below:
Return the connector to the shunt and close the door!
a)
b) From debug screen 0, Figure right, watch valve #41's 'dot'
for thirty (30) seconds. TWO (2) possible scenarios:
1)
IF (and ONLY if) valve #41's 'dot' cycles between white and blue: Turn
the machine off THEN proceed to
DIAGNOSTIC VALVE LEAK
2)
IF valve #41's 'dot' STAYS white! Call debug screen 4. Once again, watch
PDIA AND ADIA for one (1) minute! TWO (2) possible scenarios i) or ii)
below:
i)
ii)
DO NOT
press 'CONFIRM'
IF (and ONLY if) PDIA AND / OR ADIA is NOT between 2.0 and 7.5
OR changes more than 0.1 i.e. is unstable: Proceed to
462
, procedure number
IF PDIA AND ADIA is between 2.0 and 7.5 AND does NOT change
more than 0.1: Proceed to
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
TILL INSTRUCTED
page 527
TESTS.
page 527
,
SECTION 18A –
TESTS.
TMP-
3.0.0.
page 53
, procedure number
!
,
SECTION
page
F-
6.1.2.
Page 35