Fresenius Medical Care 2008K2 Troubleshooting Manual page 19

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Yes
"Operator Error" occurs! Proceed to
No
"Operator Error" DID NOT occur! See part N.
N) If done correctly, the screen says "4. Verify that 5V EST is between 4.8 to 5.2....".
screen's
O) Based on the
CHECKS:
Check #1: Is [5V EST] between 4.8 and 5.2?
Check #2: Is [12V EST] between 11.7 and 12.3?
Yes (to BOTH): Press CONFIRM to save the calibration THEN see part P.
No (to one OR both): Proceed to
END OF 'VOLTAGE DETECTOR CALIBRATION'
page 618,
[5V EST] AND [12V EST] windows, TWO (2)
page 618
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
procedure number
, procedure number
P-
H.1.2.
P-
H.1.2.
Page 15