F- 9.0.42 More Than 800 Ml Input Flow To Valve #26/Isolate Valve #24 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

F- 9.0.42 MORE THAN 800 ML INPUT FLOW TO VALVE #26 / ISOLATE VALVE #24
a) Return to Calibrate Hydraulics → Flow Pressure.
b) Press 'CONFIRM' then ENSURE the Flow Motor is rotating!
c) Remove the BLUE dialyzer connector from the shunt door.
d) Measure flow from the dialyzer connector for one (1) minute!
e) Press 'CONFIRM' to return to "Calibrate Hydraulics".
f)
More than 800 ml collected?
Yes
More than 800 ml! Good flow from Valve #24! See procedure number
No Less than 800 ml! See parts a THROUGH f below.
a)
Per the Figure below, remove Valve #24's INPUT tubing!
b)
Direct it into the bucket.
Return to Calibrate Hydraulics → Flow Pressure.
c)
d)
Press 'CONFIRM' THEN ENSURE the Flow Motor is rotating.
e)
Measure from the tubing for one (1) minute.
f)
Press 'CONFIRM' to return to "Calibrate Hydraulics".
g)
More than 800 ml collected?
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
F- 9.0.43
(page 100).