F- 1.0.1 NO WATER NEVER APPEARS / INITIAL LOADING PRESSURE CHECK
a) ENSURING the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it, SLAM* it into
the red Acetate/Acid rinse port. *OR ELSE pressure will not be read correctly!
b) Loading Pressure MAY OR MAY NOT be cycling but is it PEAKING to MORE than 15 psi?
Yes
More than 15 psi! Leaving the gauge installed, see procedure number
No Remains LESS THAN 15 psi! ENSURING the gauge was SLAMMED into the RINSE port, if
still less than 15 psi proceed to
F- 1.0.2 LOADING PRESSURE MORE THAN 15 PSI / ISOLATE "AIR IN DIALYSATE" FUNCTION
Figure below
A)
B) The female Air Sensor's connector is the
modified the connector plugs into the 'CBE board' two pins higher than the others!
C) Unplug the female Air Sensor's connector. DO NOT remove the 'CBE board' under the connector!
Leave the Air Sensor distribution board
D)
position VACANT for now!
E) Call debug screen 0.
Figure right
F)
, reading the text box above Chamber
#69, TWO (2) possible scenarios (next page):
page 116
,
remove the Distribution Board cover.
th
4
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number F- 15.0.0.
th
cap i.e. 5
position
F- 1.0.2
(page 21).
FROM THE LEFT. If CBE