AUTOTEST- 7.0.0 ISOLATE ACTUATOR-TEST BOARD / VALVES(S)
TWO (2) possibilities: 1) Bad Actuator-Test Board OR; 2) Bad valve(s) i.e. mechanically 'stuck open'.
a) Turn the machine off and swap in a known good Actuator-Test Board (see
b) Repeat the Diagnostic Valve Leak Test on the leaking valve(s). TWO (2) possible scenarios:
1)
IF passes: The previous Actuator-Test Board is bad
2) IF fails: Replace the leaking valve(s)*. *To LOCATE the valve refer to
The previous Actuator-Test Board is probably good!
Diagnostic Valve Leak Tests Theory: These Tests isolate two 'systems': System #1) During the "IOS" Test,
Air Removal Chamber #69 (IOS) and System #2) Valves for potential leaks to atmosphere:
a) Pressure is monitored by Dialysate Pressure Sensor #9.
b) Pressures are achieved during each "Prep" phase using either the UF Pump to achieve negative pressure
or Loading Pressure to achieve positive pressure. Refer to the Table below.
c) During each "Testing" phase the tested valve is closed and an alternate path to atmosphere is open for
thirty (30) seconds. If the valve is truly closed pressure holds (+/- 25 mmHg) for 30 seconds and the valve
passes.
d) During IOS all valves are closed which isolates Chamber #69 for a leak. IOS MUST pass to ensure the
valve leak tests are valid. If it fails all valves may fail!
Leak Test Pressures
Tests Requiring Negative
Pressure (approx. -300 mmHg)
Valve #43
Valve #31
Valve #33
Valve #35
Valve #37
Valve #26
Valve #24
Valve #25
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Figure
Figure 96
Tests Requiring Positive
Pressure (approx. +300 mmHg)
IOS
Valve #32
Valve #34
Valve #36
Valve #38
Valve #29
-----------
-----------
4A, page 9).
(page 534). NOTE:
Page 532