4.11.3 Substituate / dialysate / filtrate bag change

4.11.4 Ci-Ca bag change

Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Note
The bag change menu can be accessed:
– Manually by selecting the Change bag menu.
– Automatically after a bag change notification.
– Automatically after changing the procedure.
– Automatically after selecting or deselecting Ci-Ca anticoagulation.
Note
The Ci-Ca bag change menu can be accessed:
– Manually by selecting the Ci-Ca bag change menu.
– Automatically after a bag change notification.
The following information will be displayed:
The current weight on the active scale
and the required HF solution.
"Not active" (red) the deactivated
scales.
The current weight on the filtrate scales.
 Place the solution bag in question on the
scales or replace it, and / or empty the
filtrate bag.
In the event of presence of air in one of the
tubing systems, disconnect the
corresponding outlet.
 Use the rotary selector to select
Deaeration XXX, and press and hold
[OK] until all air has been removed from
the tubing system. Reconnect the
deaerated tubing system.
 Use the rotary selector to select
Terminate bag change? [OK] to
confirm! from the menu field and press
[OK].
Chapter 4: Operation
Treatment menu
4-107