Pht- 2.0.3 Filact Always = 0 - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

PHT- 2.0.3 FILACT ALWAYS = 0

TWO (2) possible scenarios:
1)
IF (and ONLY if) "Get Neg TMP" is NOT up! See procedure number
2) IF "Get Neg TMP" is still up! See parts a AND b below:
a) Figure right, remove the Fluid Sample Connector from its
port and hold it UP as shown!
b) ENSURING "Get Neg TMP" REMAINS up, allow up to
TWO (2) minutes or until you see a strong pulse that
squirts out more than six (6) feet (Yes or No)?
Yes
Strong pulse(s) seen! TWO (2) possible
scenarios:
1)
IF (and ONLY if) the "Get Neg TMP" banner is still up: Proceed to
SECTION 9 – TMP (PRESSURE)
2) ALL OTHER banners: See procedure number
No Strong pulse(s) NEVER seen! Is the "Get Neg TMP" banner still up?
Yes
"Get Neg TMP" is still up! Proceed to
PROBLEMS
No "Get Neg TMP" is NOT up! See procedure number
PROBLEMS.
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
PHT- 3.0.0
PHT- 3.0.0
(page 480).
page 503
,
SECTION 14 – UF PUMP
PHT- 3.0.0
(page 480).
(page 480).
page 440
,
Page 479