F- 1.0.5 RESET MACHINE / DEFEAT IDLE FLOW
Turn the machine OFF!
a)
Open the shunt door and leave it OPEN till instructed!
b)
c) Turn the machine on and return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!
d) From the Home screen, select the [Dialysate Flow] window.
e) Set [Dialysate Flow] to 800 ml/min and press 'CONFIRM'.
f)
Per the Figure below, check the Flow AND Deaeration Motor shafts for counterclockwise (CCW)
rotation. BOTH rotating CCW?
Yes BOTH motors rotating! See procedure number
No A motor is NOT rotating CCW! Perform parts a THROUGH c below:
a)
Close the shunt door!
b)
Call debug screen 1. ONLY if FILACT always = 0 see part c. If FILACT = 1 return to
(ABOVE) procedure number
c)
NOTING which motor is not rotating properly, proceed to
TROUBLESHOOTING MOTORS
F- 1.0.4
Figure 7 – Deaeration AND Flow Motor
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
F- 1.0.6
(page 27).
(page 24).