FIL- 6.2.0 "OPERATOR ERROR" OCCURRED / TROUBLESHOOT DIALYSATE PRESSURE
a) ENSURE the transducer protector, at the external meter, is not WET or consider replacing it!
b) Return the Dialyzer lines to the shunt and close the door.
c) Turn the machine OFF.
d) Turn the machine on and return to Dialysis Program ("Select Program" →'Dialysis' → 'CONFIRM')!
IMPORTANT!
e)
From the Home screen, select the [Dialysate Flow] window.
f)
Set [Dialysate Flow] to 800 ml/min and press 'CONFIRM'.
g) Read before performing! Return to (ABOVE) procedure number
(and ONLY if) "Operator Error" reoccurs, one at a time, swap in the listed components (see
Component List below), and in between, return to procedure number
"Operator Error" does not reoccur indicating the last component swapped in is the problem!
Component List
1) Dialysate Pressure Transducer #9
5) Sensor Board cable; 6) Distribution board.
a
To LOCATE Transducer #9 refer to
b
To prevent "Cond Offset Failure" place the machine into T and C Mode (refer to
MODES
(page Error! Bookmark not defined.)
a
; 2) Sensor Board
Figure 6
(page 19).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 437
FIL- 6.0.0
(
FIL- 6.0.0
b
; 3) Actuator-Test Board; 4) Functional Board;
) but, if
(page 437) until
OPERATING