CO- 10.0.705 ISOLATE OLC PROBLEM

IMPORTANT!
a)
TURN the machine OFF!
b) Place the concentrate connectors into their rinse ports
c) Turn the machine on and place it into Acid Clean.
d) After rinse is complete, connect to fresh concentrate and return to Dialysis Program.
e) Allow six (6) minutes for Temperature and Conductivity to stabilize.
f)
Read before performing! Return to (ABOVE) procedure number
HOWEVER, if (and ONLY if) you return here see procedure number

CO- 10.0.706 ISOLATE OLC PROBLEM

a) Read before performing! Proceed to (ABOVE) procedure number
COND CELLS HOWEVER, if (and ONLY if) you return here see part b.
b) Read before performing! Swap the following components in (see COMPONENT LIST below) one at a
time, with known good then return to (ABOVE) procedure number
new component. If (and ONLY if) you return here continue through the list until the OLC problem is
solved.
COMPONENT LIST: 1) Power Logic Board OR; 2) Actuator-Test Board OR; 3) Functional Board
1
Cond Cells MUST be calibrated after the Functional board is swapped in before its OLC
functions can be tested!
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
CO- 10.0.405
(page 385)
CO- 10.0.706
(page 393).
CO- 6.2.0
(page 354) to calibrate
CO- 10.0.405
(page 385) to test the
1