Advertisement

CO- 1.0.3 CONDUCTIVITY IS LOW / ISOLATE PUMP INPUTS
A) Plug BOTH connectors HARD into their rinse ports but DO NOT press any screen keys!
B) Figure right, two (2) checks on BOTH wands:
Check #1: ENSURE the wands are attached to the cap AND reach to the
bottom of the jug(s)!
Check #2: ENSURE the wand tubing is NOT kinked!
C) Figure right, THREE (3) checks on BOTH concentrate CONNECTORS:
Check #1:
ENSURE the O-rings are present!
Check #2:
ENSURE the filters are clean!
Check #3:
ENSURE the plug opening is not
restricted!
IMPORTANT! Place the machine in RINSE!
D)
E) ENSURING no external leaks ANYWHERE, is the external flow indicator's 'bob' moving at least ¼ up
in the sight tube?
Yes
'Bob' moving! See part F.
No
'Bob' NOT moving! Return to Dialysis Program THEN proceed to
FLOW ERRORS IN DIALYSIS PROGRAM
F) Call debug screen 1. If debug does not appear press 'Escape' then call screen 1.
G) WITHOUT LOOKING AWAY, watch NO EOS AND ALWEOS for one (1) minute!
They should NEVER = 1, not even once! Proceed per Table 7 below:
Table 6 - EOS (End Of Stroke) Values
NO EOS
0 (ALWAYS)
1 (EVER)
Does not matter
ALWEOS
0 (ALWAYS)
Proceed to
Does not matter
PUMP
Proceed to
1 (EVER)
PUMP
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
YOUR RESPONSE
See part H next page
page 397
,
SECTION 6 – CONCENTRATE
ERRORS.
page 397
,
SECTION 6 - CONCENTRATE
ERRORS.
page 20
,
SECTION 1 -