T- 1.0.2 'Bob' Moving - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

Place the machine in RINSE!
b)
c) In RINSE, is the external flow indicator's 'bob' rising at least ¼ way up in the sight tube?
Yes
No

T- 1.0.2 'BOB' MOVING

Press 'Escape' then 'CONFIRM' twice to call "Select Program" but DO NOT press
A)
'Dialysis' till instructed!
B) See procedure number
'Bob' moving! See procedure number
In RINSE, 'bob' is NOT moving! See parts a THROUGH c below:
a) Connect to acid and LIQUID bicarb.
b) RETURN TO DIALYSIS PROGRAM ("Select Program" → 'Dialysis' →
'CONFIRM')!
page 20
c) Proceed to
T- 1.0.3
(page 217).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
T- 1.0.2
(page 216).
,
SECTION 1 - FLOW ERRORS IN DIALYSIS PROGRAM
Page 216