NW- 1.0.5 "NO WATER" IN HEAT DISINFECT / ISOLATE RINSE PROGRAM
a) Press 'Escape' TWICE then 'CONFIRM' TWICE to call the "Select Program" screen!
Place the machine into RINSE (NOT HEAT DISINFECT)!
b)
c) Allow one (1) minute BEFORE continuing!
d) WITHOUT LOOKING AWAY, watch for four (4) minutes OR until if a "No Water" alarm EVER reoccurs?
Yes
"No Water" reoccurs! Proceed to
No "No Water" does NOT reoccur! The alarm may be related to Heat Disinfect ONLY! See
procedure number

NW- 1.0.6 IN RINSE "NO WATER" DOES NOT REOCCUR

a) Press 'Escape' then 'CONFIRM' twice to call the "Select Program" screen.
Place the machine in HEAT DISINFECT (NOT RINSE)!
b)
Open the shunt door and LEAVE IT OPEN until instructed!
c)
d) Figure right, from the Heat Disinfect screen,
Remaining Prerinse Time
does the TOP
window = 0:00 min:sec (Yes or No)?
Yes
Remaining Prerinse Time = 0:00! A
"No Water" alarm should not be
occurring unless the water is off!
No Remaining Prerinse Time does NOT
= 0:00! See procedure number
(page 144).

NW- 1.0.7 REMAINING PRERINSE DOES NOT = 0:00

WITHOUT LOOKING AWAY, watch for four (4) FULL minutes OR until if a "No Water" alarm EVER
reoccurs?
Yes
"No Water" alarm reoccurs! See procedure number
No
"No Water" DOES NOT reoccur! DO NOT continue but close the shunt door!
page 146
NW- 1.0.6
(page 144).
NW- 1.0.7
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
, procedure number
NW-
NW- 1.0.8
(page 145).
1.0.10.
Page 144