TMP- 6.5.0 "OPERATOR ERROR" OCCURRED / TROUBLESHOOT DIALYSATE PRESSURE
a) ENSURE the transducer protector, at the external meter, is not WET or consider replacing it!
b) Return the Dialysate lines to the shunt and close the door.
c) Turn the machine off and back on.
d) Return to Dialysis Program ("Select Program" → 'Dialysis' → 'CONFIRM')!
e) From the Home screen, select the [Dialysate Flow] window.
f)
Set [Dialysate Flow] to 500 ml/min and press 'CONFIRM'!
g) Read before performing! Repeat (ABOVE) procedure number
problem reoccurs swap in the listed components (see Component List below), one at a time, in between
returning to procedure number
last component swapped in is the problem!
Component List
1) Dialysate Pressure Transducer #9
5) Sensor Board cable; 6) Distribution board.
*
To LOCATE Transducer #9 refer to
TMP- 6.0.0
(page 470) until the problem does not occur indicating the
*
; 2) Sensor Board; 3) Actuator-Test Board; 4) Functional Board;
Figure 6
(page 19).
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
page 470
TMP- 6.0.0
(
) but if the