Fresenius Medical Care 2008K2 Troubleshooting Manual page 454

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

c) Figure below, ENSURE the Air Sensor Connectors wires terminates with the brown wire to
the TOP male probe on top of Chamber #69 AND the blue wire to the bottom probe!
IMPORTANT!
d)
Allow up to three (3) FULL minutes for FILACT to = 1. Does FILACT go
to 1?
Yes
FILACT = 1! Perform part A and B below:
IMPORTANT!
A)
to its distribution board position! If CBE equipped, using a flashlight,
ENSURE the top CBE board pin is covered by the connector.
B) Proceed to
No FILACT remains = 0! NOT LIKELY! ENSURE the Air Sensor's distribution board
position is COMPLETELY VACANT! After three (3) minutes, if (and ONLY if)
FILACT remains = 0 see procedure number
2008K/K2 Troubleshooting Guide
Return the Air Sensor's female connector PROPERLY
page 452
, procedure number
P/N 507298 Rev. H
TMP- 2.0.5
TMP- 2.0.4
(page 451).