N) Figure right, if necessary adjust the Conductivity
Limits until the 'Actual' Conductivity is CENTERED
between them.
IMPORTANT!
O)
Press 'CONFIRM'.
P) Allow up to three (3) minutes for the Conductivity
window to turn white indicating no Conductivity
alarms!
IMPORTANT! CLOSE THE SHUNT DOOR!
Q)
R) Call debug screen 2. To ENSURE the shunt door is
indicated closed CVRCLS (2
S) Call debug screen 0.
T) Allow Valve #24's 'dot' (Figure right) to turn blue BEFORE
CONTINUING to part U indicating no Temp or Cond alarms!
U) Is the flow indicator's 'bob' move at least ¼ way up in the sight
tube?
Yes
'Bob' moving! See procedure number
No
'Bob' NOT moving! See parts a THROUGH c below:
a) From the Home screen, ENSURE [Dialysate Flow] remains at 500 ml/min!
b) From debug screen 0. ENSURE Valve #24's 'dot' is REMAINING BLUE!
c) Is the external flow indicator's 'bob' rising and falling now?
Yes
No 'Bob' NOT moving! ENSURE Valve #24's 'dot' is REMAINING BLUE! If (and
nd
column from left) = 1!
'Bob' moving! See procedure number
ONLY if) 'bob' is still not moving, proceed to
ERRORS IN DIALYSIS
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
TMP- 1.0.0
(page 442).
TMP- 1.0.0
page 20
PROGRAM.
(page 442).
,
SECTION 1 - FLOW
Page 441